High-precision dual inlet system for IRMS
Iso DUAL INLET by Elementar is a high-precision dual inlet system for isotope ratio mass spectrometry (DI-IRMS), ideal for the analysis of pure gases, carbonate materials, and water samples.
Dual inlet technology is widely regarded as the most precise, accurate, and sensitive approach for isotope analysis of these sample types. Because instrumental precision directly influences data quality, the system plays a key role in applications such as paleoclimate research, where highly accurate isotope measurements are essential for reliable climate reconstructions.
Iso DUAL INLET works with:
- isoprime precisION IRMS for δ¹³C, δ¹⁵N, δ³⁴S, δ¹⁸O, and δ²H analysis
- lyticOS® Software Suite for full automation and method control
- ArDB cloud solutions for data management and quality control
Why use dual inlet technology?
Dual inlet systems allow direct comparison between sample gas and reference gas under identical conditions. This minimizes analytical bias and ensures the highest possible measurement precision.
The iso DUAL INLET uses vacuum technology to manipulate and match sample and reference gases before introducing them into the IRMS. Both gases follow identical pathways and are alternately introduced into the ion source. This enables repeated measurements of the same sample for better statistical confidence.
What are the main advantages of iso DUAL INLET?
The system is ideal to combine analytical performance with operational efficiency:
- High sensitivity for both large and very small sample volumes, including small CO₂ samples
- High precision measurements, including δ¹⁸O analysis of water with leading performance
- Compact benchtop footprint with integrated electronics and vacuum system
- High level of automation and low maintenance requirements
How does iso DUAL INLET handle different sample sizes?
The system offers a wide dynamic range, allowing analysis of both very small and large samples without adjustment.
This is achieved through:
- Two stepper motor-driven, high-compression bellows controlled automatically by software
- A patented micro-volume automatic cold finger for small sample handling
These features enable the system to analyze even minimal sample volumes with high precision and accuracy.
How is high precision achieved?
Careful control of gas handling and instrument design ensures high analytical precision.
Key elements include:
- Identical treatment of sample and reference gases through parallel flow paths
- 14 ultra-low dead volume valves machined from high-purity stainless steel
- A turbomolecular pump for the changeover valve to eliminate memory effects
This configuration enables extremely precise isotope ratio measurements, including high-precision δ¹⁸O analysis of water samples.
How does automation improve usability?
The system is fully controlled by the lyticOS® Software Suite, enabling automated operation and reducing manual intervention.
Automation includes:
- Automatic gas expansion and intensity matching
- Control of chopping routines without pre-programming
- Long unattended runs supported by large-capacity (100 ml) reference gas bellows
- Minimal maintenance due to durable valve components designed for millions of cycles
This results in high system uptime and efficient laboratory workflows.
What does the technical setup look like?
iso DUAL INLET uses a vacuum-based gas handling system to prepare samples for isotopic analysis.
Process overview:
- Turbomolecular pumps isolate and expand the sample and reference gases
- Gases are intensity-matched using variable bellows
- A changeover valve alternates gas introduction into the IRMS ion source
- The system performs multiple measurements on each sample to improve statistical accuracy
The system avoids isotopic fractionation during gas handling.
Which systems can be combined with iso DUAL INLET?
| Combined carbonate & water analysis | iso MULTI PREP | Single system for carbonate and water samples Fast switching between modes by changing needle configuration |
Mode change within minutes |
|---|---|---|---|
| High‑precision water analysis | iso AQUA PREP | δ¹⁸O water analysis using equilibration technique High stability with no measurable drift |
Precision < 0.05 ‰ (1σ, n = 10) Up to 180 samples per sequence |
| High‑sensitivity carbonate analysis | iso CARB PREP | Automated carbonate analysis down to 20 µg Dual‑core needle, direct‑drive acid pump, LN₂ cold finger for CO₂ trapping |
Up to 180 samples in heated tray |
| Automated pure gas analysis | iso MANIFOLD | Automated analysis of sealed gas samples Supports valve‑sealed tubes and break‑glass ampules |
Up to 50 valve‑sealed tubes Up to 20 break‑glass ampules |
Analytical options
iso DUAL INLET supports:
- Analysis of pure gases such as CO₂, H₂, SO₂, and O₂
- Use as a continuous flow monitoring gas injection system
- Safe handling of toxic gases by minimizing laboratory exposure
- Optional IRMS upgrades for clumped isotope analysis (Δ47)
Sample types
The system supports a wide range of sample types, depending on configuration:
- Pure gases (CO₂, O₂, H₂, SO₂)
- Groundwater, ice core samples, and biological waters
- Carbonate minerals and fossilized carbonates
- A combination of carbonate and water analyses using multi-prep configurations