Schmidt Heansch

Coloromat 100

Easy and reliable colour measurement of liquids
Measurement of transmission/extinction
Color determination of liquid samples
Up to nine fixed wavelengths
Polarimeter tubes up to 100 mm
Integrated ICUMSA methods

Coloromat 100

Spectrophotometer for colour measurement

The Coloromat 100 is a spectrophotometer enables easy and reliable colour measurement of liquids. It features a continuous measuring mode with a freely configurable rate, ideal for monitoring colour changes in product streams. Transmission, extinction, and ICUMSA liquid colour units at the standard wavelength of 420 nm are included as standard features.

 

Easy to use 

The touchscreen enables soft key functions and alphanumeric input, making it easy to enter information such as batch names and user details. Additionally, users can select wavelengths using the touchscreen. The instrument comes with three standard wavelengths and optionally integrates up to six additional custom wavelengths, enabling a broad range of applications.

Thanks to the spacious sample chamber, both polarimeter flow-through tubes up to 100 mm and standard cuvettes can be used, allowing for easy filling and high sample throughput. Moreover, using a 100 mm tube significantly improves measurement accuracy and resolution, up to ten times better than a 10 mm cuvette.

Furthermore, its sample compartment is open at both the top and bottom, preventing spills from entering the unit. With its unique optical design, the Coloromat is unaffected by stray light. This spectrophotometer also features a continuous measuring mode with a freely configurable sampling rate, ideal for monitoring colour changes in product streams over time.

 

Features of the spectrophotometer for colour measurement

  • Integrated ICUMSA methods
  • Objective and repeatable measurements
  • Up to nine fixed wavelengths
  • Polarimeter tubes up to 100 mm
  • Single or flow through measurement
  • Remote control via PC
  • Data output to PC or direct connected printer
  • Up to 250 methods programmable
  • Internal data storage of results
  • GLP/GMP conform documentation

 

Applications 

  • Enzymatic determination
  • Colour measurement of liquid crystal sugar (ICUMSA)
  • Colour index determination of beverages, e.g. soft drinks, wine, etc
  • Determination of the colour extracts (coffee, etc)
  • Colour measurement of beer and wort (EBC colour number)
  • Biochromate measurements
  • Multi-standard methods
  • Simple extinction measurements
  • Monitoring of time-dependent colour changes

 

Specifications of the spectrophotometer for colour measurement

Optics Single beam photometer
Wavelength range 340 – 900 nm*
Pre-installed filters 560, 420, 720 nm
FWHM 420 nm = 4 nm**
720 and 560 nm = 7 nm**
Wavelength selection Via touch screen
Resolution extinction 0.001
Measurement accuracy ± 0.020 extinction at 420 nm
± 0.010 extinction at 560 nm
± 0.010 extinction at 720 nm
± 0.020 extinction at optional filters

 

ICUMSA method

Measuring range 0 – 16,000 ICUMSA Units (IU)
Resolution ICUMSA 1 IU
Accuracy IU ± 10 % for colours < 20 IU
± 5 % for colours > 20 IU

 

Additional information

Light source LED
Operator Interface Touch screen for direct functions and alphanumerical inputs
Data presentation Graphic display
Data output RS232C & USB
Operating temperature 10 – 40°C
Languages English and German, others on request

* For WL < 500 nm and > 700 nm, additional LED’s may be required to ensure measurement accuracy and stability
** This very narrow half-value width enables high measurement accuracy as well as very good repeatability

Related instruments

ATR-P Touch Refractometer

High-performance robust Refractometer for the determination of mixing ratios (mass percent)
For permanent use in production laboratories
High performance and accuracy even in rugged environment

VariRef

Refractometer on the bench. Fast and precise measurements
Liquid and solid samples
Multi-point calibration for full measurement range

VariPol

The Compact Laboratory Polarimeter. The VariPol makes polarimetry faster than ever
Basic Polarimeter especially designed for pharmaceutical applications – 21 CFR Part 11 ready
Modular configuration

Last viewed products